GasAlertMicroClip
User Manual
No Gas Detected
When the detector senses a sufficient amount of
gas concentration (approximately 30 seconds), the
audible alarm beeps once,
flashes, and K
remains lit while the detector completes the span
(approximately 2 minutes).
The LCD displays FAILURE if the detector does not detect
any gas within 2 minutes before proceeding with the
calibration procedure.
Note
Successful Span
The detector will not span a sensor if:
If the sensors have spanned successfully, the audible alarm
beeps before the calibration procedure continues.
You do not apply gas to the sensor.
The sensor fails to detect at least one-half of the
expected gas concentration in the first
30 seconds.
Unsuccessful Span
If the sensors fail the span, the LCD displays FAILURE
before proceeding with the calibration procedure.
The gas concentration drops below one-half of
the expected gas level during the 2 minute span.